Fully programmable microprocessor memory
- Dual set points
- mA & VDC recorder output
- Differential input for ground loop protection
- Automatic one or two point calibration
- Last calibration data
- Manual or Automatic Temperature Compensation
- Extensive range for both conductivity and TDS
The HI 700 series of regulators offer state of the art specifications for your process control. They can be configured for ON/OFF, proportional, PI or PID control. Thanks to our exclusive technology, they can be customized to best fit your application. Bright LED’s show the current status even from a distance. A menu-driven display aids the user throughout the operations with running messages and clear prompts. All relevant parameters can be simply adjusted and will remain memorized until overwritten.
With self-diagnostic features and extractable terminals, installation and maintenance are fast and simple. Password protection guarantees that the calibration and predetermined parameters cannot be altered unnecessarily. The controllers can operate with four-ring probe or 4-20 mA signal. They accept probes with or without a built-in Pt100 temperature sensor.
Order Information:
HI 710 model is provided with dual set point, ON/OFF and PID controls and is supplied with mounting brackets and instructions.
HI 710221-2 dual setpoint, on/off and PID controls, analog output, 230V
Product Manual:
HI 710 Series (Process Instrumentation)
Specification :
| Range | Conductivity | 0.0 to 199.9 µS/cm; 0 to 1999 µS/cm 0.00 to 19.99 mS/cm; 0.0 to 199.9 mS/cm |
| TDS | 0.0 to 100.0 mg/L (ppm); 0 to 1000 mg/L (ppm) 0.00 to 10.00 g/L (ppt); 0.0 to 100.0 g/L (ppt) | |
| Temperature | -10.0 to 100.0°C | |
| Resolution | Conductivity | 0.1 µS; 1 µS; 0.01 mS; 0.1 mS |
| TDS | 0.1 ppm; 1 ppm; 0.01 g/L (ppt); 0.1 g/L (ppt) | |
| Temperature | 0.1°C | |
| Accuracy @ 20°C | EC/TDS | ±0.5% f.s. |
| Temperature | ±1°C (outside) | |
| EC Calibration | automatic or manual at 1 point | |
| Temperature Compensation |
automatic or manual, -10 to 100°C with adjustable temperature coefficient from 0.00 to 10.00%/°C | |
| TDS Conversion Factor |
adjustable from 0.00 to 1.00 | |
| Outputs | analog: isolated 0-1 mA, 0-20 mA and 4-20 mA; 0-5 VDC, 1-5 VDC and 0-10 VDC or digital: RS485 bi-directional opto-isolated | |
| Analog Input | 4-20 mA | |
| Set Point Relay | two contact outputs SPDT 5A-250 VAC, 5A-30 VDC (resistive load), fuse protected (2A, 250V fast fuse) | |
| Alarm Relay | contact output SPDT 5A-250 VAC, 5A-30 VDC (resistive load), fuse protected (2A, 250V fast fuse) | |
| Power Supply | 115 VAC ±10% or 230 VAC ±10%; 50/60 Hz | |
| Power Consumption | 15 VA | |
| Over Current Protection |
400 mA 250V fast fuse | |
| Environment | 0 to 50°C (32 to 122°F); RH max 95% non-condensing | |
| Dimensions | panel cutout: 140 x 140 mm, instrument: 144 x 144 x 170 mm | |
| Weight | 1.6 kg (3.5 lb.) | |
| Dimensions | 81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5”) |
HI713 Checker HC ® - Phosphate, LR
HI723 Checker HC ® - Chromium VI, HR
HI708 Checker HC ® - Nitrite, HR
HI5522 Research Grade pH/ORP/ISE and EC/TDS/Salinity/Resistivity Meter with CAL Check™
เครื่องวัดปริมาณคลอรีนรวม รุ่น HI711
HI701 Checker HC ® - Free Chlorine
BL983320-1 EC Mini Controller 0 to 199.9 µS/cm, 230V
เครื่องควบคุมค่าความนำไฟฟ้า
เครื่องวัดปริมาณโบรมีน ในตัวอย่างน้ำ รุ่น HI716
HI38058 CTK pH
HI7033L - EC 84 µS/cm@25°C - 500ml
HI70031C - EC 1413 µS/cm@25°C Zert. - 25x20ml
HI7032L - 1382 mg/L (ppm) TDS @25°C - 500ml
HI739 Checker HC ® - Fluoride. HR
HI705 Checker HC ® - Silica, LR
HI38020 CTK Chlorine Free & Total
เครื่องวัดปริมาณแอมโมเนีย ช่วงต่ำ รุ่น HI700
เครื่องวัดปริมาณคลอรีนอิสระ รุ่น HI701
BL983318-1 TDS Mini Controller 0.0 to 10.00 g/L (ppt) - 230V 










