- Easy to handle
- Fire retardant casing
- Selectable overdose protection system
- Splash resistant cover
BL 983314 is a simple to operate resistivity controller designed for ultra pure water, reverse osmosis, and water conditioning applications. The BL 983314 resistivity controller is also ideal for continuous monitoring of process solutions. Set point and calibration are manually adjusted with a trimmer and the alarm relay allows for simple control.
Readings are automatically temperature compensated, with three different coefficients (β=2.4, 3.5 or 4.5 %/°C). The alarm contact can be used for connection to an alarm, pump, solenoid or dosing system.
The relay contact is open when readings are higher than the set point, while for measurements lower than set point, the relay contact is closed. The Hysteresis is typically 0.20 MW/cm from the set point.
Measurements are displayed on the LCD and the multicolored LED continuously indicates if the controller is in measurement, dosing or alarm mode. Users can choose automatic or manual dosing mode with a switch on the front panel. Manual control is particularly useful during maintenance operations, because it permits operators to enable or disable the dosing relay according to need.
The overtime control system advises users when the relay is active for too long, helping to prevent over-dosage.
Order Information:
BL 983314-1 ( 115/230V) is supplied with mounting brackets, transparent cover and instruction manual.
Manual: Download
Specification :
| Range | 0.00 to 19.90 MΩ/cm | |
| Resolution | 0.10 MΩ/cm | |
| Accuracy @ 25°C | ±2% F.S. | |
| Probe | HI 3314 resistivity probe with 2 m (6.6’) cable (included) | |
| Temperature Compensation | Automatic and linear, 5 to 50°C (41 to 122°F) | |
| Temperature Coefficient | ß=2.4; 3.5; 4.5%/°C selectable through jumper on the rear panel |
|
| Calibration | factory calibrated | |
| Dosing Relay | maximum 2A (fuse protected), 250 Vac, 30 Vdc contact close when measure < set point | |
| Setpoint | adjustable from 0 to 19.90 MΩ/cm | |
| Overtime | adjustable, typically from 5 to approx. 30 minutes | |
| Power Supply | See Chart | |
| Dimensions | 79 x 49 x 95 mm (3.1 x 1.9 x 3.7’’) |
HI93751-01 Sulfate - Reagents - 100 Tests
HI983302N EC Indicator up to 9.99 mS with Electrode GRO'CHEK
เครื่องวัดปริมาณฟอสเฟต ช่วงต่ำ รุ่น HI713
เครื่องวัดปริมาณคลอรีนอิสระ รุ่น HI701
เครื่องวัดปริมาณฟอสเฟต ช่วงสูง รุ่น HI717
Reagents for HI 717 Checker HC ®, 40 Tests
HI38000 CTK Sulfate
เครื่องวัดปริมาณโบรมีน ในตัวอย่างน้ำ รุ่น HI716
HI73127 Spare-pH-Electrode for Tester
HI93708-01 Nitrite HR - Reagents - 100 Tests
HI700 Checker HC ® - Ammonia, LR
BL983321-1 TDS Mini Controller 0.00 to 19.99 mg/L (ppm) - 230V
HI93704-01 Hydrazine - Reagents - 100 Tests
HI7072 Electrolyte Solution, 1M KNO³, 4 x 30 ml
HI4013-03 ISE 1000 mg/L (ppm) Nitrate Std , 500 ml Bottle
HI5003 - 3.00 pH@25°C - TQ - 500ml
BL983322-1 EC Mini Controller, 0 to 19.99 µS/cm, 230V 