HI8936 คือเครื่องส่งสัญญาณค่าการนำไฟฟ้าที่ออกแบบใหม่ของเรา ซึ่งใช้โพรบโพเทนชิโอเมตริกแบบสี่วงแหวน โพรบนี้แทบจะต้านทานการปนเปื้อนจากสารละลายที่ไม่สะอาดได้ ซึ่งช่วยให้เครื่องส่งทำงานด้วยประสิทธิภาพสูงสุดตลอดเวลา
- Utilizes HANNA’s potentiometric four-ring conductivity probe
- Automatic Temperature Compensation
- Direct probe connection eliminates signal loss
- Use in conjunction with HI 7638 or HI 3001 platinum conductivity probes
HI8936 is our redesigned conductivity transmitter that utilizes a four-ring potentiometric probe. This probe is virtually immune to contamination by unclean solutions. This allows the transmitter to operate at peak performance at all times.
Temperature effects are compensated for by utilizing both the built-in temperature sensor on the probe and the transmitter’s ATC circuitry with a β of 2%/°C.
Direct connection of the probe to the transmitter assures a positive electrical connection with no signal loss over long distances.
HI 8936”L” versions allow easy verification and monitoring of measured values and is easier to calibrate and maintain.
The HI 8936 series requires external power to the 4-20 mA current loop.
Order Information:
All HI 8936 models are supplied complete with instructions.
Product Manual:
HI 8936BN (Process Instrumentation)
Specification :
| Range | 0.00 to 19.99 mS/cm | |
| Accuracy @ 20°C | ±2% f.s. (excluding probe error) | |
| Calibration | manual, two point, with offset and slope trimmers | |
| Temperature Compensation | fixed or automatic from 0 to 50°C (32 to 122°F) with ß=2%/°C | |
| Conductivity Probe | HI 7635 for in-line applications (not included) | |
| Recorder Output | 4-20 mA, not isolated, max 500 Ohm | |
| Protection | IP65 | |
| Power Supply | 12-30 VDC | |
| Environment | 0 to 50°C (32 to 122°F); RH max 95% non-condensing | |
| Dimensions | 165 x 110 x 71 mm (6.5 x 4.3 x 2.8’’) | |
| Weight | 1 kg (2.2 lb.) | 
 HI93719-01 Hardness Magnesium - Reagents -100 Tests
HI93719-01 Hardness Magnesium - Reagents -100 Tests						 HI93703-05 AMCO-AEPA-1 Calibration Solution at 500 FTU, 30 ml-Copy
HI93703-05 AMCO-AEPA-1 Calibration Solution at 500 FTU, 30 ml-Copy						 Reagents for HI 715 Checker HC ®, 25 Tests
Reagents for HI 715 Checker HC ®, 25 Tests						 HI7091L - Reducing Pretreatment Solution, 500 ml
HI7091L - Reducing Pretreatment Solution, 500 ml						 HI93102 Water Analysis Meter (with Turbidity up to 50NTU)
HI93102 Water Analysis Meter (with Turbidity up to 50NTU)						 สารละลายมาตรฐานความขุ่น 20 NTU
สารละลายมาตรฐานความขุ่น 20 NTU						 HI7072 Electrolyte Solution, 1M KNO³, 4 x 30 ml
HI7072 Electrolyte Solution, 1M KNO³, 4 x 30 ml						 HI96802 Fructose - Digital Refractometer Portable
HI96802 Fructose - Digital Refractometer Portable						 HI731335 - Cap for Cuvette 4 pcs. - Serie HI967xx
HI731335 - Cap for Cuvette 4 pcs. - Serie HI967xx						 เครื่องวัดคลอรีน
เครื่องวัดคลอรีน						


 
				 
				 
				 
				 
				
 
				